QuantStudio qPCR
Revolution drives two QuantStudio instruments, and the integration depth is very different between them:
| Device | Revolution control |
|---|---|
| QuantStudio 7 Pro | Full run lifecycle — queue, start, pause, resume, stop, retrieve results, drawer control, power state |
| QuantStudio 7 | Tray in and out, and data file parsing |
The specifications below are Thermo Fisher’s for the QuantStudio 6 and 7 Pro systems.
At a glance
Section titled “At a glance”| Temperature range | 4–99.9 °C |
| Temperature uniformity | 0.4 °C |
| Maximum ramp rate | Up to 9.0 °C/sec (96-well 0.1 mL Fast block) |
| Block formats | 96-well 0.2 mL, 96-well 0.1 mL (Fast), 384-well, TaqMan Array Cards |
| Dynamic range | 10 orders of magnitude |
| Filter / colour combinations | 21 on the QuantStudio 7 Pro (5 on the 6 Pro) |
| Run time | Less than 30 min |
Specifications
Section titled “Specifications”Thermal
Section titled “Thermal”Temperature range, uniformity and run time are already given in the At a glance table above. Beyond that:
| Heating and cooling | Peltier |
| VeriFlex blocks | 3 or 6 independent temperature zones (96-well blocks) |
| Block size | 22.5 × 22.5 × 7 cm |
Maximum ramp rate, by block:
| Block | Maximum | Average sample |
|---|---|---|
| 96-well (0.2 mL) | 6.5 °C/sec | 3.66 °C/sec |
| 96-well (0.1 mL) | 9.0 °C/sec | 4.81 °C/sec |
| 384-well | 6.0 °C/sec | 2.92 °C/sec |
| TaqMan Array Cards | 5.0 °C/sec | 3.00 °C/sec |
VeriFlex is worth knowing about: on 96-well blocks it divides the block into 3 or 6 independent temperature zones, so different cycling conditions can run in the same plate. That is an optimisation tool, and it means “the block temperature” may not be a single value.
Reaction volumes
Section titled “Reaction volumes”| Block | Volume |
|---|---|
| 96-well 0.1 mL | 10–30 µL |
| 96-well 0.2 mL | 10–100 µL |
| 384-well | 5–20 µL |
| TaqMan Array Cards | ~1.5 µL |
Blocks are interchangeable and tool-free to swap, guided by the onboard software — but Thermo Fisher notes the system ships with one block type and others are bought separately, and that blocks are released at intervals so some may not currently be available.
Optics and detection
Section titled “Optics and detection”Filter/colour combinations and dynamic range are already given in the At a glance table above.
| Excitation | Bright white LED |
| Excitation / emission range | 450–600 nm / 500–640 nm |
| Detection | CMOS |
| Data acquisition | Whole-plate imaging |
| Multiplexing | 5 or 6 targets |
| Sensitivity | 1.5-fold resolution; detect down to 1 copy |
Dye compatibility as listed: FAM, SYBR Green I, VIC, JOE, HEX, TET, ABY, NED, TAMRA, Cy3, JUN, ROX, Texas Red, Mustang Purple, Cy5, LIZ and Cy5.5.
System
Section titled “System”| Dimensions (D × W × H) | 52.5 × 33.8 × 54.7 cm |
| Weight | 38 kg |
| Onboard memory | 10 GB, approximately 450 run files |
| Touchscreen | 12 inch capacitive, with real-time application viewing |
| Communication | Connect Platform, USB, LAN or Wi-Fi |
| External devices | 2D barcode reader via USB |
| Configuration | Stand-alone, PC connected, or LAN / Wi-Fi to the Connect Platform |
| Electrical approvals | IEC, CE |
| Calibration | Engineer calibrated during installation |
| Data export | RDML, for MIQE compliance |
Supported methods — QuantStudio 7 Pro
Section titled “Supported methods — QuantStudio 7 Pro”Running
Section titled “Running”- Start run / Stop run — begin and abandon a run
- Run template — run from a stored template
- Pause run — pause, holding at a given temperature
- Resume run — resume a paused run
Pause at a temperature is the operation that matters most here. A qPCR run interrupted without a hold temperature is a compromised run; pausing at a defined temperature is what makes an intervention mid-protocol survivable. Thermo Fisher lists programmable and manual pause among the system’s run-programming options.
Run queue
Section titled “Run queue”- Send to run queue — add a run to the queue
- List run queue — read what is queued
- Remove from run queue — take a run off the queue
The queue is what lets a schedule stage work without waiting for the block. Runs are under 30 minutes each, so a cell that loads plates faster than the cycler processes them needs somewhere to put them.
Results
Section titled “Results”- List completed runs — enumerate what is on the instrument
- Download completed run — retrieve a run
- Export results — export result data
- Delete completed run — remove a run by name
With 10 GB of onboard memory holding roughly 450 run files, retrieval and deletion are the pair that keeps an unattended instrument from filling up.
Instrument state
Section titled “Instrument state”- Open drawer / Close drawer — present the drawer for a robot
- Get plate and block state — read whether a plate is present and which block is fitted
- Power up / Stand by — bring the instrument up, or put it into standby
Get plate and block state is the check to make before committing a transfer: it answers both “is the drawer empty” and “is the right block in there”, and the block determines the reaction volume the plate must match.
Supported methods — QuantStudio 7
Section titled “Supported methods — QuantStudio 7”- Extend tray / Retract tray — present and withdraw the tray for a robot
- Parse data file — parse a results data file
This is a plate-access and data-parsing integration rather than run control. Runs are started on the instrument.
Integration
Section titled “Integration”The 7 Pro is reached over the network — IP address and port, or a URL — authenticated with an access key, with working and results folders on the host. The older QuantStudio 7 is addressed by instrument name.
Prerequisites
Section titled “Prerequisites”- The instrument must be reachable, and the access key valid, before the device is initialised.
- Working folder and results folder must exist and be writable.
- The block fitted must match the plate the method presents — check with get plate and block state rather than assuming.
- Protocol templates must already exist on the instrument or in Thermo Fisher’s software.
Device configuration
Section titled “Device configuration”| Property | Purpose |
|---|---|
| IP Address / Instrument Port | Address of the 7 Pro |
| URL | Service endpoint |
| Access Key | Credential for the instrument API |
| Working Folder | Working directory used during a run |
| Results Folder | Where retrieved results are written |
| Instrument Name | Identifies the older QuantStudio 7 |
If you need more of this instrument driven from a schedule, get in touch — the driver is extended on demand.
Operations
These are available from a Run Device Operation step.
Close Drawer
Closes the instrument's plate drawer, under the same interlocks that
Delete Completed Run
Deletes a finished run from the instrument's run history, freeing the
| Parameter | Type | Default | Description |
|---|---|---|---|
| Run Name | string | — | Name of the finished run to delete from the instrument, as |
Download Completed Run
Copies a finished run's results file off the instrument to the machine
| Parameter | Type | Default | Description |
|---|---|---|---|
| Run Name | string | — | Name of the finished run to download, as published by the |
| Destination Folder | string | — | Folder to write the downloaded results file into, resolved on the |
Export Results
Converts a downloaded results file into report files, written as CSV
| Parameter | Type | Default | Description |
|---|---|---|---|
| EDS File Path | string | — | Full path to the downloaded results file to export, resolved on |
| Destination Folder | string | — | Folder to write the exported files into, resolved on the machine |
| File Name Prefix | string | — | Prefix for the names of the exported files. The results file's own |
| Format | string | — | Export file format: CSV, XLS, XLSX, TXT, ZIP or RDML, matched |
| As Single File | bool | — | Whether to write one combined file instead of a separate file for |
Get Instrument Info
Reports the instrument's identity, serial numbers, firmware and
Get Plate And Block State
Reports whether a sample block is fitted to the instrument and whether
Get Run Progress
Reports the active run's name and how far it has reached through the
Get Status
Reports what the instrument is doing now: its state, drawer position,
List Completed Runs
Reports how many finished runs the instrument still holds in its run
List Run Queue
Reports how many templates are waiting in the instrument's run queue,
Open Drawer
Opens the instrument's plate drawer, so that a plate can be placed on
Pause Run
Pauses the run in progress, holding the sample block at the temperature
| Parameter | Type | Default | Description |
|---|---|---|---|
| Temperature Celsius | double | — | Temperature in degrees Celsius at which the sample block is held |
Power Up
Brings the instrument out of standby and back to an idle, run-ready
Raise Error
Makes this device report a failure on demand: the step fails with an error naming the device, and the operator is offered the usual error recovery choices. No command reaches the instrument and nothing moves, and the failure is raised in simulation as well as on a real system, so this is the step to use when rehearsing how a schedule handles a device failure.
Remove From Run Queue
Withdraws a waiting template from the instrument's run queue,
| Parameter | Type | Default | Description |
|---|---|---|---|
| Template File Name | string | — | Name of the queued template to withdraw, as it appears in the run |
Resume Run
Continues a run that was paused, taking the protocol up from the point
Run Template
Starts a qPCR run from an EDT template file, waits for the instrument
| Parameter | Type | Default | Description |
|---|---|---|---|
| Template File Path | string | — | Full path to the EDT run template that defines the protocol and |
| Run Name | string | — | Name that identifies this run on the instrument and names the |
| Plate Barcode | string | — | Barcode to record in the run's plate setup. When one is supplied |
| Timeout Seconds | int | — | How long to wait for the whole run, in seconds, counting the lamp |
| Download Results | bool | — | Whether to download the results file once the run finishes. The |
Send To Run Queue
Uploads an EDT template file to the instrument's own run queue without
| Parameter | Type | Default | Description |
|---|---|---|---|
| Template File Path | string | — | Full path to the EDT run template to upload to the instrument's |
Stand By
Puts the instrument into its low-power standby state, which turns the
Start Run
Starts a qPCR run from an EDT template file and returns as soon as the
| Parameter | Type | Default | Description |
|---|---|---|---|
| Template File Path | string | — | Full path to the EDT run template that defines the protocol and |
| Run Name | string | — | Name that identifies this run on the instrument and names the |
| Plate Barcode | string | — | Barcode to record in the run's plate setup. When one is supplied |
Stop Run
Aborts the run in progress. The run ends where it stands and cannot be